9 Diagnostics and service
VEGAFLEX 83 • Modbus and Levelmaster protocol
41838-EN-130314
Fault description Error pattern Cause Rectication
5. Measured value
jumps sporadically to
100 % during lling
Level
time
0
Varying condensation or con-
tamination on the probe
Carry out false signal suppres-
sion
6. Measured value
jumps to ≥ 100 % or
0 m distance
Level
time
0
Level echo is no longer
detected in the close range
due to false signals in the close
range. The sensor goes into
overll protection mode. The
max. level (0 m distance) as
well as the status message
"Overll protection" are output-
ted.
Eliminate false signals in the
close range
Check installation conditions
If possible, switch o the func-
tion "Overll protection"
Measurement error during emptying
Fault description Error pattern Cause Rectication
7. Measured value re-
mains unchanged in
the close range during
emptying
Level
False echo larger than the level
echo
Level echo too small
Eliminate false signals in the
close range
Remove contamination on the
probe. After having removed
the source of the false signals,
the false signal suppression
must be deleted.
Carry out a new false signal
suppression
8. Measured value re-
mains reproducible
in one position during
emptying
Level
time
0
Stored false signals in this
position are larger than the
level echo
Delete false signal memory
Carry out a new false signal
suppression
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Setup" must be carried out again or must
be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no.
+49 1805 858550.
The hotline is also available outside the normal working hours on
seven days a week around the clock.
Since we oer this service worldwide, the support is in the English
language. The service itself is free of charge, the only costs involved
are the normal call charges.
9.5 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user.
In Ex applications only one instrument and one electronics module
with respective Ex approval may be used.
If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron-
Reaction after fault recti-
cation
24 hour ser
vice hotline