9 Diagnostics and service
VEGAFLEX 83 • Modbus and Levelmaster protocol
41838-EN-130314
1
2
Level
time
0
Fig. 25: The broken line 1 shows the real level, the continuous line 2 shows the
level displayed by the sensor
Note:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value"
In case of a too low level indication, the reason could be a line
resistance that is too high
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value
shows a too low or too
high level
Level
time
0
Min./max. adjustment not
correct
Adapt min./max. adjustment
Wrong linearization curve Adapt linearization curve
Running time error (small
measurement error close to
100 %/serious error close to
0 %)
Repeat setup
2. Measured value
jumps towards 100 %
Level
time
0
Due to the process, the ampli-
tude of the product echo sinks
A false signal suppression was
not carried out
Carry out false signal suppres-
sion
Amplitude or position of a
false signal has changed (e.g.
buildup); false signal suppres-
sion no longer matches
Determine the reason for the
changed false signals, carry out
false signal suppression, e.g.
with buildup
Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3. Measured value re-
mains in the area of the
bottom during lling
Level
time
0
Echo from the probe end larger
than the product echo, for
example, with products with
ε
r
< 2.5 oil-based, solvents, etc.
Check application parameter
"Medium" and "Vessel height",
adapt if necessary
4. Measured value re-
mains momentarily
unchanged during ll-
ing and then jumps to
the correct level
Level
0
Turbulence on the product
surface, quick lling
Check application parameters,
change if necessary, e.g. in
dosing vessel, reactor