38
6 Set up with the display and adjustment module
VEGAFLEX 83 • Modbus and Levelmaster protocol
41838-EN-130314
The instrument carries out an automatic false signal suppression
as soon as the probe is uncovered. The false signal suppression is
always updated.
The menu item "Delete" is used to completely delete an already cre-
ated false signal suppression. This is useful if the saved false signal
suppression no longer matches the metrological conditions in the
vessel.
In the menu item "Lock/unlock adjustment", you can protect the
sensor parameters against unauthorized modication. The PIN is
activated/deactivated permanently.
The following adjustment functions are possible without entering the
PIN:
Select menu items and show data
Read data from the sensor into the display and adjustment mod-
ule.
Caution:
With active PIN, adjustment via PACTware/DTM as well as other
systems is also blocked.
You can change the PIN number under "Additional adjustments -
PIN".
This menu item enables the setting of the requested national lan-
guage.
In the delivery status, the sensor is set to the ordered national lan-
guage.
In this menu item, you dene the indication of the measured value
on the display. You can display two dierent measured values. In this
menu item, you dene measured value 1.
The default setting for the displayed value 1 is "Filling height Level".
Lock/unlock setup - Ad-
justment
Display - Menu language
Display - Displayed value
1