79
11 Anexo
VEGAFLEX 83 • Protocolo Modbus y Levelmaster
41838-ES-130416
Register Name Register Number Type Note
TV 2106 DWord
ThirdVariableinByteOrderABCDDCBA(LittleEn-
dian)
QV 2108 DWord
QuarternaryVariableinByteOrderDCBA(LittleEn-
dian)
Status 2200 DWord See Register 100
PV 2202 DWord
PrimaryVariableinByteOrderBACD(MiddleEndian)
SV 2204 DWord SecondaryVariableinByteOrderBACD(MiddleEn-
dian)
TV 2206 DWord
ThirdVariableinByteOrderBACD(MiddleEndian)
QV 2208 DWord QuarternaryVariableinByteOrderBACD(MiddleEn-
dian)
Unit Codes for Register 104, 108, 112, 116
Unit Code Measurement Unit
32 Degree Celsius
33 Degree Fahrenheit
40 US Gallon
41 Liters
42 Imperial Gallons
43 Cubic Meters
44 Feet
45 Meters
46 Barrels
47 Inches
48 Centimeters
49 Millimeters
111 Cubic Yards
112 Cubic Feet
113 Cubic Inches
11.4 ModbusinstruccionesRTU
FC3 Read Holding Register
Con esa instrucción se puede leer una cantidad arbitraria (1-127) en los registros holding. Se trans-
miten el registro inicial, desde el que hay que leer y la cantidad de registros.
Request:
Parameter Length Code/Data
Function Code 1 Byte 0x03
Start Address 2 Bytes 0x0000 to 0xFFFF