Vega VEGABAR 64 4 … 20 mA_HART Manual de usuario Pagina 51

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 80
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 50
Connect the sensor to power supply and provide connection from PC
to the
instrument via VEGACONNECT. Start PA CTware and provide
connection to the sensor, e.g. via the VEGA project assistant. Close
the parameter window of the sensor, as far as open.
Go in the PACTware menu bar to "Instrument data", "Additional
functions" and "Update instrument software".
PACTware checks now the actual hardware and software version of
the sensor and displays the data. This procedure lasts approx. 60 s.
Push the button " Update software" and select the previously
extracted hex le. Then the software update can be started. Further
les are installed automatically. Depending on the sensor, this
procedure lasts approximately 1 h.
8.6 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Dow
nloads - Forms and certicates - Repair
form".
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
l Print and ll out one form per instrument
l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can nd the respective agency on our website
www.vega.com
unde
r: "Compa
ny - VEGA worldwide"
Prepare update
Load software into sen-
sor
VEGABAR 64 4 20 mA/HART 51
8 Maintenance and fault recticati
on
27525-EN-081209