Level and Pressure
Operating Instructions
Hydrostatic pressure
transmitters D85
Vista de pagina 0
1 2 3 4 5 6 ... 43 44

Indice de contenidos

Pagina 1 - Operating Instructions

pLevel and PressureOperating InstructionsHydrostatic pressuretransmitters D85

Pagina 2 - Note Ex area

10 Pressure transmitter D85Electrical dataAdjustment rangesZero adjustable from -20 … +95 % of nominal rangeSpan adjustable from 3.3 … 120 % of nomina

Pagina 3 - Pressure transmitter D85 3

Pressure transmitter D85 11Product descriptionSupply and signal circuit (analogue transmission, 4 … 20 mA),additional data for electronics version LSu

Pagina 4 - 1 Product description

12 Pressure transmitter D85Protective measures 1)Protection IP 66, IP 67, IP 68VEGABOX 01 IP 66 and IP 67Protection class IIIOvervoltage category IIIA

Pagina 5 - Pressure transmitter D85 5

Pressure transmitter D85 13Product descriptionOperating conditionsAmbient conditionsAmbient temperature -40°C … +85°C- with indicating module -10°C …

Pagina 6 - Electronics version I:

14 Pressure transmitter D851.7 DimensionsHousingsProduct descriptionø 7667ø 7591ø 41,64663ø 41,693ø 41,64690

Pagina 7 - , VVO adjustable

Pressure transmitter D85 15ø 41,6SW 6025G 1½ Aø 55GG135 (174*)ø 41,6SW 60251½ " NPTGN135 (174*)ø 41,6SW 4621M44 x 1,25BA135 (174*)ø 41,6ø 64CA135

Pagina 8 - Electronics version L:

16 Pressure transmitter D85Product descriptionPressure transmitter accessoryWelded socketø 6519GG21ø 60BARd 65 x 1/6ø 79,640LAø 72/6xM6ø 79,640LB21,5ø

Pagina 9 - 1.5 Technical data

Pressure transmitter D85 17Product description8510811813538 72ø 5M20x1,513013972100with protective coverVEGABOX 01

Pagina 10 - 18 24 30 36

18 Pressure transmitter D85Mounting2 Mounting2.1 Mounting instructionsThe pressure transmitter can be mounted inany position. Cable entries must point

Pagina 11 - Supply voltage

Pressure transmitter D85 19Electrical connection3 Electrical connection3.1 Connection instructionsThe electronics in the pressure transmitterrequires

Pagina 12

2 Pressure transmitter D85ContentsContentsSafety information ... 2Note Ex area ..

Pagina 13 - Operating conditions

20 Pressure transmitter D85Electronics version K4 … 20 mA output signal with integratedadjustment and remote parameter settingthrough PC and HART® han

Pagina 14 - 1.7 Dimensions

Pressure transmitter D85 21Electrical connection3.3 Connection to external connection housing VEGABOX 01Zum Anschluss anDruckmessumformer mitanalogem

Pagina 15 - Process connections D85

22 Pressure transmitter D85Electrical connectionSupply through a VEGA signal conditioning instrumentStandard wiring for non-standardised output.Supply

Pagina 16 - Welded socket

Pressure transmitter D85 23Setup4 Setup4.1 Adjustment structureThe hydrostatic pressure transmitters comewith or without adjustment capability.After c

Pagina 17 - VEGABOX 01

24 Pressure transmitter D85Adjustment with PC and adjustmentprogram VVO (electronics version B, I, Kand L)With the adjustment program VVO (VEGAVisual

Pagina 18 - 2 Mounting

Pressure transmitter D85 25SetupAdjustmentFor adjustment of zero and span, an amme-ter must be connected to terminals 1 and 3.The measured value is id

Pagina 19 - 3 Electrical connection

26 Pressure transmitter D85 250 Ri < 250 +-PLCSetup4.4 Setup with HART® handheldSensors with electronics versions I, K and Lcan be set up with th

Pagina 20 - Electronics version B

Pressure transmitter D85 27SetupThe most important adjustment stepsOn the following four pages you see a menuschematic of the HART® handheld in conjun

Pagina 21 - Electronics version L

28 Pressure transmitter D85SetupHART® menu schematic123451.1 4.1(5.1) 4.2(5.2)Less importantmenu windowsNot required, unimportantor blocked menu windo

Pagina 22 - + -

Pressure transmitter D85 29SetupBlocked menu. Thefollowing menu window isindicated but not supportedby the sensor. Thereforesaving of adjustmentscarri

Pagina 23 - 4.1 Adjustment structure

Pressure transmitter D85 33 Electrical connection3.1 Connection instructions ... 193.2 Terminal assi

Pagina 24 - 24 Pressure transmitter D85

30 Pressure transmitter D85SetupHART® menu schematic (continuation)1.3.31.3.46 Write protect None7 Descriptor8 Message9 PV Snsr s/nFinal asmbly num 0R

Pagina 25 - Integration time

Pressure transmitter D85 31Setup1.2.3.1.11.2.3.1.2Note:After entering a parameter, push "ENTER" and then "SEND".Confirm the messag

Pagina 26 - handheld

32 Pressure transmitter D852221) Ri < 250 4 ... 20 mA Setup5.5 Adjustment with PC directly on the sensor(electronics versions I, K and L)For conne

Pagina 27 - Pressure transmitter D85 27

Pressure transmitter D85 33SetupIf you have already connected the PC withthe adjustment software VVO to your measur-ing system,• first switch on the p

Pagina 28 - menu schematic

34 Pressure transmitter D85• Confirm the adjustment with " " and aftersaving you are again in the initial menu.SetupNote:If you connect the

Pagina 29 - Pressure transmitter D85 29

Pressure transmitter D85 35You can carry out the min./max. adjustmentwith or without medium. Generally you willcarry out the adjustment without medium

Pagina 30 - 30 Pressure transmitter D85

36 Pressure transmitter D85• Click to " ".• When the sensor is unpressurised and ininstallation position, confirm the safetyenquiry with &qu

Pagina 31 - Pressure transmitter D85 31

Pressure transmitter D85 37You can choose four linearisation curves:linear, horiz. cylindrical tank, spherical tankand user programmable curve.Under &

Pagina 32

38 Pressure transmitter D85Max. 32 index markers (value pairs) can beentered.• Quit the menu with " ".• Confirm the message with " &quo

Pagina 33 - User access

Pressure transmitter D85 39Saved sensor data can be transferred lateron to other sensors.For example, if you have a system with sev-eral of the same s

Pagina 34 - Configuration/Measuring

4 Pressure transmitter D85Product description1 Product description1.1 Function and configurationThe pressure transmitter D85 is an efficientinstrument

Pagina 35 - Pressure transmitter D85 35

40 Pressure transmitter D85Diagnostics6 DiagnosticsMaintenanceThe pressure transmitter D85 is maintenancefree.Should dismounting of the pressure trans

Pagina 36 - Parameter adjustment

Pressure transmitter D85 41Instrument modification7 Instrument modification7.1 Retrofitting the adjustment insertSuch a retrofitting can be necessary,

Pagina 37 - Offset correction

42 Pressure transmitter D85Notes

Pagina 38 - 38 Pressure transmitter D85

Pressure transmitter D85 43Notes

Pagina 39 - Meas. data info

ISO 9001All statements concerning scope of delivery, application, practicaluse and operating conditions of the sensors and processing sys-tems corresp

Pagina 40 - 6 Diagnostics

Pressure transmitter D85 5Product description1.2 Electronics version without adjustmentElectronics version B:Pressure transmitter for connection to VE

Pagina 41 - 7 Instrument modification

6 Pressure transmitter D85Product descriptionElectronics version I:Pressure transmitter 4 … 20 mA, HART®, VVOVVO512344-20mAterminalKlemmeinsatz+-12V..

Pagina 42 - 42 Pressure transmitter D85

Pressure transmitter D85 71.3 Electronics version with integrated adjustment in connection housingElectronics version K:Pressure transmitter 4 … 20 mA

Pagina 43 - Pressure transmitter D85 43

8 Pressure transmitter D851.4 Electronics for connection to VEGADIS 12VEGADIS 12DISPLAY+ -TRANSMITTER5876321101112OPERATEPOINTDISPLAYTRANSMITTERENDZE

Pagina 44 - 2.24 987 / January 2001

Pressure transmitter D85 9Product description1.5 Technical dataMechanical dataMaterials, wetted partsProcess connection stainless steel 1.4571, 1.4435